An FPGA-based design for power efficient low delay rate adaptive pacemaker using accelerometer and heart rate sensor
Rohini Srivastava1, Ch Kalyan Kumar Prusty1, Nitin Sahai2
1Electronics and Communication Engineering Department, Motilal Nehru National Institute of Technology Allahabad, Prayagraj, India.
Abstract:
This paper presents a power efficient, low delay and rate adaptive dual chamber pacemaker (PLRDPM) using heart rate and accelerometer sensor. In recent years, number of modifications have been done in the pacemaker design. However, design of an implantable device on an open source is still challenging. Through this paper, we are proposing a "proof of concept" for the design of PLRDPM on FPGA for improving the vital parameters: delay and power consumption. The proposed PLRDPM comprises of accelerometer and heart rate sensors to measure physical activity's effect on heart rate of the bradycardia patients. A rate adaptive pacing algorithm has been designed using two sensor's data, to reduce the delay and power consumption. However, delay in the responses of various components in the circuitry produces an accumulative delay effect in any practical circuit. The delay and the power consumption for the proposed PLRDPM are found to be 2.82 ns and 9 mW, respectively.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Special considerations while measuring pulse