Differences in partial oxygen saturation obtained from three pulse oximeters
Luis Efrén Santos-Martínez1, Martha Vianey Ureña-Plascencia2, Javier Quevedo-Paredes2
1Instituto Mexicano del Seguro Social, Centro Médico Nacional Siglo XXI, Hospital de Cardiología, Departamento de Hipertensión Pulmonar y Corazón Derecho.
Pulse oximeter variability was assessed in 255 patients. Onyx II and Hergom devices showed low variability, while the Rossmax SB220 exhibited significant variability, making it unsuitable for clinical use.
Area of Science:
- Medical Devices
- Clinical Measurement
- Respiratory Medicine
Background:
- Variability in partial oxygen saturation measurements exists between different pulse oximeters.
- Accurate oxygen saturation monitoring is crucial for patient care.
Purpose of the Study:
- To evaluate and compare the measurement variability of three distinct pulse oximeters in a clinical setting.
- To determine the reliability of different pulse oximeter devices for patient monitoring.
Main Methods:
- A cross-sectional study involving 255 consecutive patients.
- Demographic, occupational, and disease data were collected.
- Variability was assessed using Bland-Altman analysis and intraclass correlation coefficient (ICC).
Main Results:
- The study included patients with a mean age of 61 years, predominantly female and smokers, with COPD being a common condition.
- Onyx II and Hergom pulse oximeters demonstrated low measurement variability (ICC > 0.949).
- The Rossmax SB220 showed substantial variability (ICC = 0.784), indicating poor agreement.
Conclusions:
- Onyx II and Hergom pulse oximeters are reliable for clinical use due to their low variability.
- The Rossmax SB220 is not recommended for clinical practice owing to its high measurement variability.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
12:29Magnetic Resonance Imaging Quantification of Pulmonary Perfusion using Calibrated Arterial Spin Labeling
Published on: May 30, 2011
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring pulse