Hygrometric Performances of Different High-Flow Nasal Cannula Devices: Bench Evaluation and Clinical Tolerance
Mathieu Delorme1, Pierre-Alexandre Bouchard1, Serge Simard1
1Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec, Canada. Mr Delorme is affiliated with the Université Paris-Saclay, UVSQ, ERPHAN,Versailles, France.
Background:
High-flow nasal cannula (HFNC) is increasingly used for the management of respiratory failure. Settings include [Formula: see text], total gas flow, and temperature target. Resulting absolute humidity (AH) at the nasal cannula may affect clinical tolerance, and optimal settings with respect to hygrometry remain poorly documented.
Methods:
A bench study was designed to assess AH delivered by 4 HFNC devices (Optiflow, Airvo 2, Precision Flow, and Hydrate) according to flow, ambient temperature, and other available settings. Clinical tolerance of different levels of hygrometry (20, 30, and 40 mg H2O/L) was evaluated in 15 healthy volunteers.
Results:
With [Formula: see text] set at 1.0, normal ambient temperature, and settings made accordingly to the manufacturers' recommendations, mean ± SD AH was 42.2 ± 3.1, 39.5 ± 1.8, 35.7 ± 2.0, and 32.9 ± 2.7 mg H2O/L for the Airvo 2, Optiflow, Hydrate, and Precision Flow, respectively, (P < .001). AH dropped from -3.5 to -10.7 mg H2O/L (P <. 001) with high ambient temperature, except for the Precision Flow. Increasing flow did not significantly affect AH except for the Precision Flow (from 36.4 ± 1.6 to 29.8 ± 0.2 mg H2O/L at 10 and 40 L/min, respectively, [P < .001]). The lowest AH was encountered with the Optiflow set with noninvasive ventilation (NIV) mode, without compensation algorithm, and at high ambient temperature (14.2 ± 1.5 mg H2O/L). In studied subjects, AH significantly affected breathing comfort, reduced from 7.0 ± 1.0 to 3.0 ± 2.0 at 40 and 20 mg H2O/L, respectively, (P < .001). Comfort was similar at 30 and 40 mg H2O/L.
Conclusions:
When used according to manufacturer's recommendations and at normal ambient temperature, all the HFNC devices evaluated achieved satisfactory hygrometric output with respect to breathing comfort evaluated in healthy subjects (≥ 30 mg H2O/L). Substantial differences exist between devices, and optimal knowledge of their working principles is required as inappropriate usage may dramatically alter efficacy and clinical tolerance.
More Related Videos
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...