Related Experiment Video
Updated: Oct 19, 2025

Remote Laboratory Management: Respiratory Virus Diagnostics
Published on: April 6, 2019
Study on contaminant distribution in a mobile BSL-4 laboratory based on multi-region directional airflow
Yan Wang1, Jian-Tao Miao2, Jian-Bo Chen3
1School of Mechanical Engineering, University of Shanghai for Science and Technology, Shanghai, 200093, China. yanwang909909@163.com.
Abstract:
The outbreak of COVID-19 has caused increasing public attention to laboratory-acquired infections (LAIs), especially for a mobile Bio-Safety Level 4 Lab (BSL-4) with high potential of exposure. In this paper, the distribution and removal mechanism of bioaerosols in the biosafety laboratory were studied. A simulation model of airflow distribution in the opening and closing state of air-tight door was established and verified. The results showed that the airflow entrainment velocity during the opening of the door was approximately 0.12 m/s. It increased the probability of vortex generation in the laboratory. The deposition rate of particles was doubled when the air-tight door opening is compared with air-tight door closing. Besides, nearly 80% of the particles deposited on the surface of the wall and ceiling, increasing the possibility of LAIs. The findings of this paper could provide new scientific methods for high-level biosafety laboratories to avoid cross-infection. Moreover, future work regarding air-tight door rotation speed regulation and control should be emphasized.