Related Experiment Video
Updated: Oct 19, 2025

A Novel Technique of Rescuing Capsulorhexis Radial Tear-out using a Cystotome
Published on: January 16, 2011
Evaluation of ophthalmic surgical simulators for continuous curvilinear capsulorhexis training
Nilesh Raval1, Vivian Hawn, Mimi Kim
1From the Department of Ophthalmology and Visual Sciences, Montefiore Medical Center, Bronx, New York (Raval, Shrivastava); Albert Einstein College of Medicine, Bronx, New York (Hawn); Department of Epidemiology & Population Health, Albert Einstein College of Medicine, Bronx, New York (Kim, Xie).
Purpose:
To evaluate performance and assessments by expert surgeons tasked to create a continuous curvilinear capsulorhexis (CCC) on 3 commercially available surgical simulators.
Setting:
Montefiore Medical Center Department of Ophthalmology and Visual Sciences, Bronx, New York.
Design:
Randomized, cross-sectional, comparative study.
Methods:
Expert cataract surgeons (N = 7) were tasked to create a 5.5-mm CCC on 3 surgical simulators (Bioniko, Kitaro, and SimulEYE). Surgeons rated how well each simulator approximated human tissue on a modified Likert scale (1 to 7). Duration, size, and number of forceps grabs were evaluated.
Results:
7 surgeons performed a total of 63 trials. Bioniko required a greater number (6.53 ± 3.14) of forceps grabs for CCC creation than Kitaro (4.90 ± 2.47, P = .01) and SimulEYE (3.90 ± 1.34, P < .0001). Surgeons created the 5.5-mm CCC most accurately on Bioniko and SimulEYE, with the largest mean CCC performed on Kitaro (8.00 ± 0.84) compared with that on Bioniko (5.24 ± 0.60, P < .0001) and SimulEYE (5.11 ± 0.41, P < .0001). Surgeons spent more time (seconds) performing the CCC on Bioniko (41.95 ± 26.70) than that on Kitaro (32.05 ± 14.99, P = .02) and SimulEYE (28.90 ± 15.18, P = .002). Kitaro (4.56 ± 0.84, P < .0001) and SimulEYE (4.19 ± 0.92, P < .0001) were rated as more realistic than Bioniko (1.38 ± 0.80).
Conclusions:
SimulEYE and Kitaro were believed to most closely approximate human capsular tissue, and surgeons performed the CCC fastest on these models. However, surgeons created a 5.5-mm CCC most accurately on SimulEYE and Bioniko. SimulEYE had the best overall performance and fidelity across all studied metrics; however, each simulator demonstrated its own unique advantages and disadvantages. Larger validation studies will help residency programs best use training tools for novice surgeons.