Variation in Dosimeter Calibration Factor (NDW) Over a Period of 20 Years

Manoj Kumar Semwal1, Harishchandra Gupta2, Sakshi Singhal2

  • 1Department of Radiation Oncology, Army Hospital Research and Referral, Delhi, India.

Journal of Medical Physics
|September 27, 2021
PubMed
Summary

COVID-19 pandemic caused delays in calibrating radiation dosimeters, raising concerns about patient dose accuracy. A 20-year review found the absorbed dose to water calibration factor (NDW) for ion chambers remained stable, < ±3% change.

Related Concept Videos

Biological Effects of Radiation02:59

Biological Effects of Radiation

All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they...
16.3K
Instrument Calibration01:12

Instrument Calibration

Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
348
Dose Size and Dosing Frequency: Determination Methods01:21

Dose Size and Dosing Frequency: Determination Methods

Determining the optimal dose size and dosing frequency in pharmacotherapy is crucial for achieving therapeutic effectiveness while minimizing adverse effects. This article explores the methodologies employed in determining these parameters, focusing on their significance and interplay to tailor dosing regimens.Dose Size: Dose size refers to the amount of a drug administered in a single dose. It is determined based on the drug's pharmacodynamics and pharmacokinetics properties and...
23
Standard Deviation of Calculated Results01:14

Standard Deviation of Calculated Results

Standard deviation measures the spread of data around the mean value. Many large data sets follow a Gaussian distribution, also known as a normal distribution. This distribution is bell-shaped curved, with the most frequently observed value (mean or central value) in the middle. The farther away from the central value, the greater the deviation from the central value, and the lower the frequency.
A broad Gaussian distribution curve has a wider standard deviation, representing a data set with...
7.4K
Calibration Curves: Correlation Coefficient01:10

Calibration Curves: Correlation Coefficient

In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the...
3.2K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.4K