Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Novel phantom for performance evaluation of contrast-enhanced 3D rotational angiography
L E Lubis1, R A Basith2, I Hariyati3
1Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok 16424, Indonesia.
Purpose:
This technical note presents an in-house phantom with a specially designed contrast-object module constructed to address the need for three-dimensional rotational angiography (3DRA) testing.
Methods:
The initial part of the study was a brief evaluation on the commercially available phantom used for 3DRA and computed tomography angiography (CTA) to confirm the need for a special phantom for 3D angiography. Once confirmed, an in-house phantom was constructed. The novel phantom was tested to evaluate the basic image performance metrics, i.e., unsharpness (MTF) and noise characterization (NPS), as well as to show its capability for vessel contrast visibility study.
Results:
The low contrast objects in the commercially available tools dedicated for CT is found to yield significantly lower signal difference to noise ratio (SDNR) when used for 3DRA, therefore deemed inadequate for 3DRA contrast evaluation. The constructed in-house phantom demonstrates a capability to serve for basic imaging performance check (MTF, NPS, and low contrast evaluation) for 3DRA and CTA. With higher and potentially adjustable visibility of contrast objects as artificial vessels, the in-house phantom also makes more clinically relevant tests, e.g., human- or model observer study and task-based optimization, possible.
Conclusion:
The novel phantom with special contrast object module shows higher visibility in 3DRA compared to the currently available commercial phantom and, therefore, is recommended for use in 3D angiography.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT