Related Experiment Video
Updated: Oct 18, 2025

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Fricke gel xylenol orange dosimeter layers for stereotactic radiosurgery: A preliminary approach
P Pérez1, P Rico Torres2, A Bruna3
1Laboratorio de Investigación e Instrumentación en Física Aplicada a la Medicina e Imágenes de Rayos X (LIIFAMIR(x)), Facultad de Matemática, Astronomía, Física y Computación, Universidad Nacional de Córdoba, Córdoba, Argentina; Instituto de Física Enrique Gaviola (IFEG), CONICET, Córdoba, Argentina.
Abstract:
Investigations regarding the feasibility, reliability, and accuracy of Fricke gel dosimeter layers for stereotactic radiosurgery are presented. A representative radiosurgery plan consisting of two targets has been investigated. Absorbed dose distributions measured using radiochromic films and gelatin Fricke Gel dosimetry in layers have been compared with dose distributions calculated by using a treatment planning system and Monte Carlo simulations. The different dose distributions have been compared by means of the gamma index demonstrating that gelatin Fricke gel dosimeter layers showed agreements of 100%, 100%, and 93%, with dose and distance tolerances of 2% and 2 mm, with respect to film dosimetry, treatment planning system and Monte Carlo simulations, respectively. The capability of the developed system for three-dimensional dose mapping was shown, obtaining promising results when compared with well-established dosimetry methods. The obtained results support the viability of Fricke gel dosimeter layers analyzed by optical methods for stereotactic radiosurgery.