Radiographic quality control measurements comparing D-speed film, E-speed film, and xeroradiography
Abstract:
This study evaluated quality control (QC) measurements of D-speed film, E-speed film, and xeroradiography under clinical conditions. Automatic and manual (wet-tank) film-processing techniques were assessed. The results of QC tests (speed index, contrast index, and base + fog measurements) indicated that D-speed film and E-speed film react similarly under clinical conditions with regard to speed and contrast when processed automatically. E-speed film had an elevated base + fog measurement over D-speed film in both processing methods. Manually processed D-speed film had a greater contrast than did manually processed E-speed film. Xeroradiography maintained a consistent level of image processing under clinical conditions during this 90-day trial.