A selective and inert stationary phase combining triptycene with tocopheryl polyethylene glycol succinate for
1Key Laboratory of Cluster Science, Ministry of Education of China, Beijing Key Laboratory of Photoelectronic/Electrophotonic Conversion Materials, School of Chemistry and Chemical Engineering, Beijing Institute of Technology, Beijing 100081, China.
Abstract:
This work reports a selective and inert triptycene-based stationary phase (TPT) combining the triptycene framework with tocopheryl polyethylene glycol succinate (TPGS) units for capillary gas chromatography (GC). The TPT stationary phase was physically coated onto a capillary column by static coating method with the column efficiency of 4200 plates/m and moderate polarity. As demonstrated, the TPT column exhibited high inertness towards organic bases, including basic heterocycles, aliphatic and aromatic amines, showing distinct advantages over the TPGS and commercial columns. Also, the TPT column displayed high-resolution performance towards the isomers of methylpyridines, toluidines, xylidines and alkanes (C6-C8). Moreover, it showed excellent separation repeatability and reproducibility with RSD values in the range of 0.03%-0.07% for run-to-run, 0.12%-0.18% for day-to-day and 2.3%-3.6% for column-to- column (n = 4). Its applications to purity test of chemical products and to GC-MS analysis of the essential oil of Artemisia annua L. demonstrated its good potential for practical analyses. The present work has novelty in constructing highly selective and inert stationary phases and providing a feasible strategy for concurrently addressing the related problems in GC analyses. Its methodology and findings is of important value in terms of fundamental researches and practical applications.
More Related Videos
11:44Cellular Membrane Affinity Chromatography Columns to Identify Specialized Plant Metabolites Interacting with Immobilized Tropomyosin Kinase Receptor B
Published on: January 19, 2022
11:08The Use of Gas Chromatography to Analyze Compositional Changes of Fatty Acids in Rat Liver Tissue during Pregnancy
Published on: March 13, 2014
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...