Validation of a nonwearable device in healthy adults with normal and short sleep durations
David A Hsiou1,2, Chenlu Gao3,4, Robert C Matlock5
1Department of Psychology and Neuroscience, Baylor University, Waco, Texas.
Study Objectives:
To determine the accuracy of early and newer versions of a nonwearable sleep tracking device relative to polysomnography and actigraphy, under conditions of normal and restricted sleep duration.
Methods:
Participants were 35 healthy adults (mean age = 18.97; standard deviation = 0.95 years; 77.14% female; 42.86% White). In a controlled sleep laboratory environment, we randomly assigned participants to go to bed at 10:30 pm (normal sleep) or 1:30 am (restricted sleep), setting lights-on at 7:00 am. Sleep was measured using polysomnography, wristband actigraphy (the Philips Respironics Actiwatch Spectrum Plus), self-report, and an early or newer version of a nonwearable device that uses a sensor strip to measure movement, heart rate, and breathing (the Apple, Inc. Beddit). We tested accuracy against polysomnography for total sleep time, sleep efficiency, sleep onset latency, and wake after sleep onset.
Results:
The early version of the nonwearable device (Beddit 3.0) displayed poor reliability (intraclass correlation coefficient [ICC] < 0.30). However, the newer nonwearable device (Beddit 3.5) yielded excellent reliability with polysomnography for total sleep time (ICC = 0.998) and sleep efficiency (ICC = 0.98) across normal and restricted sleep conditions. Agreement was also excellent for the notoriously difficult metrics of sleep onset latency (ICC = 0.92) and wake after sleep onset (ICC = 0.92). This nonwearable device significantly outperformed clinical-grade actigraphy (ICC between 0.44 and 0.96) and self-reported sleep measures (ICC < 0.75).
Conclusions:
A nonwearable device showed better agreement than actigraphy with polysomnography outcome measures. Future work is needed to test the validity of this device in clinical populations.
Citation:
Hsiou DA, Gao C, Matlock RC, Scullin MK. Validation of a nonwearable device in healthy adults with normal and short sleep durations. J Clin Sleep Med. 2022;18(3):751-757.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
08:36Collecting Sleep, Circadian, Fatigue, and Performance Data in Complex Operational Environments
Published on: August 8, 2019
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Sleep-Wake Cycles
NREM Sleep
NREM sleep comprises four progressive stages that seamlessly merge:
Insufficient Sleep and Sleep Deprivation
Sleep deprivation is a more severe form of sleep loss...
Sleep Apnea
The condition is more prevalent among...
Substance Use Disorders Affecting Sleep
Understanding the concepts of physical dependence,...