Related Experiment Video
Updated: Oct 17, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Algorithm for an automatic treatment planning system using a single-arc VMAT for prostate cancer.
Takumi Kodama1,2, Shigehiro Kudo1, Shogo Hatanaka2
1Department of Radiation Oncology, Ina, Saitama Prefectural Hospital Organization Saitama Cancer Center, Saitama, Japan.
An automatic treatment planning system was developed for single-arc volumetric-modulated arc therapy (VMAT) in prostate cancer. This innovation reduces plan variability and improves efficiency while maintaining high-quality radiation dose distributions.
Area of Science:
- Radiation Oncology
- Medical Physics
- Cancer Treatment
Background:
- Intensity-modulated radiation therapy (IMRT) planning shows significant variability due to manual optimization.
- Existing automatic toolkits for VMAT in prostate cancer often require multiple arcs, complicating clinical implementation.
- Single-arc VMAT is preferred for prostate cancer to manage intrafraction motion and improve accuracy.
Purpose of the Study:
- To develop an automatic treatment planning system utilizing a single arc for prostate cancer VMAT.
- To enhance efficiency and reduce plan quality variations in radiation therapy.
- To ensure accurate dose delivery for prostate cancer patients.
Main Methods:
- A novel algorithm was designed for the Monaco treatment planning system to generate single-arc VMAT plans.
- The system optimizes plans by first determining cost function parameters and then iterating calculations until dose indices are met.
- Plan quality was evaluated by comparing automated plans against manually generated plans.
Main Results:
- The automatic system successfully generated single-arc VMAT plans for all patients within a few iterations.
- No statistically significant differences in plan quality were observed between automatic and manual planning for targets and organs at risk.
- The system required minimal human input, primarily initial template setup and initiation.
Conclusions:
- The developed automatic single-arc VMAT planning system is clinically suitable for prostate cancer.
- This automated approach significantly improves efficiency and resource utilization in treatment planning.
- The system consistently delivers high-quality dose distributions, reducing inter-planner variability.