Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
A single-step, high throughput, and highly reproducible method for measuring IgM quantity and avidity directly from
Angus Li1, Richard J Harris1, Bryan G Fry1
1The University of Queensland, School of Biological Sciences, Brisbane, QLD, 4072, Australia.
Abstract:
Quantification of specific antibody responses is critical in determining activation of MHCII-dependent immune memory and is generally performed by enzyme-linked immunosorbent assay (ELISA). Antibody avidity for a particular antigen is also informative of the quality of the adaptive immune response following vaccination. Avidity can be determined by chaotropic elution ELISA, pre-absorption ELISA, or surface plasmon resonance (SPR), although multimeric antibodies such as IgM are problematic for SPR. ELISA-based assays are very time consuming, require secondary antibody reagents, and are poorly repeatable. Here we demonstrate that biolayer interferometry (BLI) using an Octet HTX instrument can robustly and reproducibly quantify and determine avidity of specific IgM for an antigen directly from fish serum in a single step. We collected sera from giant grouper (Epinephelus lanceolatus) that had been vaccinated with the hapten 2,4-dinitrophenol conjugated to keyhole limpet hemocyanin (DNP-KLH) and from control fish injected with phosphate buffered saline. The specific IgM in the serum and its avidity for DNP were quantified via ELISA and BLI. BLI was precise and highly repeatable for determination of the quantity and avidity of antibody in the serum compared to ELISA. The wet-lab preparation and machine running time for BLI was 3-5 times faster than ELISA to generate the same amount of data. The ELISA inter-plate variation significantly affected reproducibility while BLI was consistent and repeatable between samples and plates. Indeed, the consistency of BLI data indicated that technical triplicates were redundant. Biological replication alone was sufficient to elucidate the effect of treatments. However, BLI required a lower serum dilution than ELISA for similar sensitivity, and thus more serum was required to produce high resolution data. BLI is an extremely high-throughput assay, providing teleost serum IgM quantification and avidity data as a single-step, agile alternative to ELISA.
More Related Videos
07:08A Rapid, Multiplex Dual Reporter IgG and IgM SARS-CoV-2 Neutralization Assay for a Multiplexed Bead-Based Flow Analysis System
Published on: April 6, 2021
08:52Use of an Influenza Antigen Microarray to Measure the Breadth of Serum Antibodies Across Virus Subtypes
Published on: July 26, 2019