Does the location of electrode on tympanic membrane matter when recording electrocochleography?
Mohd Normani Zakaria1, Nik Adilah Nik Othman2, Zalilah Musa3
1Audiology Programme, School of Health Sciences, Universiti Sains Malaysia, Kubang Kerian, Malaysia.
Background:
The non-invasive tympanic electrocochleography (TM-ECochG) is useful for clinical diagnoses. Nevertheless, the influence of the electrode location on tympanic membrane (TM) on ECochG results needs to be studied.
Objective:
The aim of the present study was to compare the TM-ECochG results obtained when the electrode was placed on the superior region versus the inferior region of TM.
Materials And Methods:
Forty healthy adults (aged 29 to 50 years) participated in this comparative study. The TM-ECochG testing was conducted with the electrode placed on the superior and inferior regions of TM.
Results:
SP and AP amplitudes were statistically higher for the inferior region of TM (p < .05). In contrast, SP/AP ratios were comparable between the two regions of TM (p = .417).
Conclusions And Significance:
In TM-ECochG recording, when the electrode was placed on the inferior region of TM, SP and AP amplitudes were greater than when the electrode was placed on the superior region of TM. On the other hand, SP/AP amplitude ratio was not affected by the location of electrode on TM. The findings from the present study could be useful to guide clinicians in optimizing TM-ECochG recording when testing their respective patients.