In vitro CT visualization of the biopsy location by filling the core biopsy needle with contrast media
Related Experiment Video
Updated: Aug 5, 2026

Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
Published on: April 14, 2011
Per Thunswärd1, David Eksell1, Håkan Ahlström1
1Department of Surgical Sciences - Radiology, 8097Uppsala University, Uppsala, Sweden.
Background:
When performing computed tomography (CT)-guided biopsy procedures with non-disposable, automatic biopsy instruments, the actual course of the biopsy needle is not registered.
Purpose:
To evaluate the ability to visualize the sampling location after CT-guided biopsy in vitro using a novel method, where the space between the inner needle and the outer cannula in a core biopsy needle is filled with contrast media; and to compare the grade of visibility for two different concentrations of contrast media.
Material And Methods:
Core needle biopsies were performed in a tissue phantom using biopsy needles primed with two different iodine contrast media concentrations (140 mg I/mL and 400 mg I/mL). Commercially available needle-filling contraptions with sealing membranes were used to fill the needles. Each biopsy was imaged with CT, and the visibility was evaluated twice by three senior radiologists in a randomized order.
Results:
The presence of traces was confirmed after biopsy, almost without exception for both concentrations. The visibility was sufficient to determine the biopsy location in all observations with the 400 mg I/mL filling, and in 7/10 observations with the 140 mg I/mL filling. The grade of visibility of the trace and the proportion of the biopsy needle course outlined were higher with the 400 mg I/mL filling.
Conclusion:
With CT-guided biopsy in vitro, the sampling location can be visualized using a novel method of priming the biopsy needle with iodine contrast media, specifically highly concentrated contrast media.