2D RF pulse design for optimized reduced field-of-view imaging at 1.5T and 3T
Orhun Caner Eren1, Bahadir Alp Barlas2, Emine Ulku Saritas3
1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey; Institute of Neuroinformatics, ETH Zurich, Zurich, Switzerland; Department of Information Technology and Electrical Engineering, ETH Zurich, Zurich, Switzerland.
Abstract:
Two-dimensional spatially selective radiofrequency (2DRF) excitation pulses are widely used for reduced field-of-view (FOV) targeted high-resolution diffusion weighted imaging (DWI), especially for anatomically small regions such as the spinal cord and prostate. The reduction in FOV achieved by 2DRF pulses significantly improve the in-plane off-resonance artifacts in single-shot echo planar imaging (ss-EPI). However, long durations of 2DRF pulses create a sensitivity to through-plane off-resonance effects, especially at 3T where the off-resonance field doubles with respect to 1.5T. This work proposes a parameter-based optimization approach to design 2DRF pulses with blips along the slice-select axis, with the goal of maximizing slab sharpness while minimizing off-resonance effects on 1.5T and 3T MRI scanners, separately. Extensive Bloch simulations are performed to evaluate the off-resonance robustness of 2DRF pulses. Three different metrics are proposed to quantify the similarity between the actual and ideal 2D excitation profiles, based on the signals within and outside the targeted reduced-FOV region. In addition, simulations on a digital brain phantom are performed for visual comparison purposes. The results show that maintaining a sharp profile is the primary design requirement at 1.5T, necessitating the usage of relatively high slab sharpness with a time-bandwidth product (TBW) around 8-10. In contrast, off-resonance robustness is the primary design requirement at 3T, requiring the usage of a moderate slap sharpness with TBW around 5-7.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....