Clinical and Dosimetric Implications of Calculating Lung Shunt Fraction for Hepatic 90Y Radioembolization Using
Lucas Struycken1, Mikin Patel1, Phillip Kuo1
1Department of Medical Imaging, University of Arizona, 1501 N Campbell Ave, Tucson, AZ 85724.
Abstract:
BACKGROUND. Accurate assessment of hepatopulmonary shunting, typically performed by planar scintigraphy, is critical in planning 90Y radioembolization. High lung shunt fractions (LSFs) may alter treatment. OBJECTIVE. The purpose of this study is to compare LSFs calculated from planar scintigraphy versus SPECT/CT in patients with high planar LSFs (> 15%) and to describe the potential clinical and dosimetric implications of SPECT/CT LSF calculations. METHODS. This retrospective study included 36 patients (29 men and seven women; mean age, 62.4 ± 9.8 [SD] years) who underwent 99mTc-macroaggregated albumin (MAA) planar scintigraphy for planning hepatic radioembolization, had a planar LSF greater than 15%, and underwent concurrent SPECT/CT. Clinically reported planar LSFs were recorded. SPECT/CT LSFs were retrospectively calculated using automatically generated volumetric ROIs around the lungs and liver with subsequent manual adjustments. Total lung and perfused liver doses were calculated using a medical internal radiation dose model. Values derived from planar and SPECT/CT data were compared using Mann-Whitney U tests. Multivariable regression analysis was performed of factors associated with the discrepancy in LSF between the techniques. RESULTS. Mean planar LSF was 25.1% ± 11.6%, and mean SPECT/CT LSF was 16.0% ± 9.3% (p < .001). Mean lung dose was 18.8 ± 8.0 Gy for planar LSF versus 12.3 ± 7.2 Gy for SPECT/CT LSF (p < .001). Mean perfused liver dose was 92.9 ± 36.1 Gy using planar LSF versus 102.7 ± 39.1 Gy using SPECT/CT LSF (p < .001). In multivariable analysis, a larger discrepancy in LSF between planar scintigraphy and SPECT/CT was associated with a body mass index (weight in kilograms divided by the square of height in meters) of 26 or higher (p = .02), maximum tumor size of less than 9 cm (p = .05), and left hepatic intraarterial injection (p = .02). Fourteen of 36 patients did not undergo upfront radioembolization due to a planar LSF greater than 20% and instead underwent shunt-reducing embolization with subsequent radioembolization (n = 7), transarterial chemoembolization (n = 5), or no treatment (n = 2). Five of these 14 patients had a SPECT/CT LSF of less than 20% and would have been eligible for upfront radioembolization based on SPECT/CT LSF. Seven of 29 patients treated with radioembolization underwent prescribed dose reductions based on planar LSF; six of these patients would have qualified for standard radioembolization without dose reduction using SPECT/CT LSF. CONCLUSION. Planar scintigraphy yields greater LSFs compared with SPECT/CT, possibly leading to unnecessary shunt-reducing procedures and prescribed dose reductions. CLINICAL IMPACT. SPECT/CT should be considered for clinical LSF calculations before radioembolization in patients with high LSFs.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...