Related Experiment Video
Updated: Oct 15, 2025

Use of Capillary Electrophoresis Immunoassay to Search for Potential Biomarkers of Amyotrophic Lateral Sclerosis in Human Platelets
Published on: February 10, 2020
PC Deficiency Testing: Thrombin-Thrombomodulin as PC Activator and Aptamer-Based Enzyme Capturing Increase Diagnostic
Sara Reda1, Heiko Rühl1, Jana Witkowski1
1Institute of Experimental Hematology and Transfusion Medicine, University Hospital Bonn, Bonn, Germany.
Improving protein C (PC) activity testing accuracy is crucial for diagnosing thrombophilia. Replacing the Protac activator with thrombin-thrombomodulin (TM) significantly enhances diagnostic sensitivity for PC deficiency detection.
Area of Science:
- Biochemistry
- Hematology
- Molecular Diagnostics
Background:
- Protein C (PC) activity assays are standard for detecting PC deficiency in thrombophilia workups.
- Current methods use Protac activation and chromogenic or clot-based detection, with variable diagnostic accuracy.
- Investigating alternative activation and detection methods is essential for improving diagnostic sensitivity.
Purpose of the Study:
- To evaluate the impact of different PC activation methods (Protac vs. thrombin-TM) and APC detection techniques on diagnostic accuracy.
- To compare chromogenic, clot-based, and aptamer-based assays for PC activity measurement.
- To identify optimal strategies for sensitive and accurate PC deficiency diagnosis.
Main Methods:
- Prospective cohort study of 31 patients with confirmed PC deficiency.
- Comparison of PC activation using Protac versus thrombin-thrombomodulin (TM) complex.
- Assessment of activated PC (APC) detection via chromogenic assay, clot-based assay, and an oligonucleotide-based aptamer assay.
Main Results:
- Chromogenic and clot-based assays identified 84% and 77% of PC deficient patients, respectively.
- The aptamer-based assay increased true-positive results to 87%.
- Replacing Protac with thrombin-TM for activation yielded true-positive results in all patients (100% sensitivity).
Conclusions:
- The method of PC activation critically influences the accuracy of PC activity testing.
- Thrombin-TM complex activation significantly improves diagnostic sensitivity compared to Protac.
- Aptamer-based APC detection offers high sensitivity, especially for mutations outside the active site, and is less prone to interference than clot-based assays.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
03:38Aptamer-Based Target Detection Facilitated by a 3-Stage G-Quadruplex Isothermal Exponential Amplification Reaction
Published on: October 6, 2022