Quantitative SPECT (QSPECT) at high count rates with contemporary SPECT/CT systems
Alessandro Desy1,2, Guillaume F Bouvet1,2, Étienne Croteau3,4
1Department of Medical Imaging, and Research Centre (Oncology Axis), CHU de Québec - Université Laval, 11 côte du Palais, Quebec City, QC, G1R 2J6, Canada.
Background:
Accurate QSPECT is crucial in dosimetry-based, personalized radiopharmaceutical therapy with 177Lu and other radionuclides. We compared the quantitative performance of three NaI(Tl)-crystal SPECT/CT systems equipped with low-energy high-resolution collimators from two vendors (Siemens Symbia T6; GE Discovery 670 and NM/CT 870 DR).
Methods:
Using up to 14 GBq of 99mTc in planar mode, we determined the calibration factor and dead-time constant under the assumption that these systems have a paralyzable behaviour. We monitored their response when one or both detectors were activated. QSPECT capability was validated by SPECT/CT imaging of a customized NEMA phantom containing up to 17 GBq of 99mTc. Acquisitions were reconstructed with a third-party ordered subset expectation maximization algorithm.
Results:
The Siemens system had a higher calibration factor (100.0 cps/MBq) and a lower dead-time constant (0.49 μs) than those from GE (75.4-87.5 cps/MBq; 1.74 μs). Activities of up to 3.3 vs. 2.3-2.7 GBq, respectively, were quantifiable by QSPECT before the observed count rate plateaued or decreased. When used in single-detector mode, the QSPECT capability of the former system increased to 5.1 GBq, whereas that of the latter two systems remained independent of the detectors activation mode.
Conclusion:
Despite similar hardware, SPECT/CT systems' response can significantly differ at high count rate, which impacts their QSPECT capability in a post-therapeutic setting.
Related Concept Videos
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
NMR Spectrometers: Overview
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...