Related Experiment Video
Updated: Oct 14, 2025

Dual-mode Imaging of Cutaneous Tissue Oxygenation and Vascular Function
Published on: December 8, 2010
Wearable Personal Core Body Temperature Measurement Considering Individual Differences and Dynamic Tissue Blood
Abstract:
Accurate and continuous measurement of the human core body temperature by a wearable device is of great significance for human health care and disease monitoring. The current wearable thermometers ignore the physiological differences between individuals and the role of blood perfusion in thermoregulation, resulting in insufficient accuracy and limitations in terms of the measurement sites. This study proposed a novel personal model for measuring core body temperature by taking dynamic tissue blood perfusion and individual differences into consideration. The technique facilitates possible accurate core body temperature measurements from the skin surface of the wrist and forehead. First, the personal core body temperature model was established based on the thermal equilibrium between the human body and the measurement device, in which the tissue blood perfusion changes dynamically with tissue temperature. Then, the parameters of the personal model that imply individual physiological differences were obtained based on personal data collected daily. The results show that with the developed personal model, the accuracy of the measured body temperature from the wrist is close to that of the forehead model. The wrist model and the forehead model have a mean absolute error of 0.297 (SD = 0.078)°C and 0.224 (SD = 0.071)°C, respectively, which meets the accuracy and robustness requirements of practical applications. The personal models significantly improve the accuracy compared with that of the group model, especially for the wrist model.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...