Validation of Heart Rate Extracted From Wrist-Based Photoplethysmography in the Perioperative Setting: Prospective
Eveline Mestrom1, Ruben Deneer2,3,4, Alberto G Bonomi5
1Department of Anesthesiology, Catharina Hospital Eindhoven, Eindhoven, Netherlands.
Background:
Measurement of heart rate (HR) through an unobtrusive, wrist-worn optical HR monitor (OHRM) could enable earlier recognition of patient deterioration in low acuity settings and enable timely intervention.
Objective:
The goal of this study was to assess the agreement between the HR extracted from the OHRM and the gold standard 5-lead electrocardiogram (ECG) connected to a patient monitor during surgery and in the recovery period.
Methods:
In patients undergoing surgery requiring anesthesia, the HR reported by the patient monitor's ECG module was recorded and stored simultaneously with the photopletysmography (PPG) from the OHRM attached to the patient's wrist. The agreement between the HR reported by the patient's monitor and the HR extracted from the OHRM's PPG signal was assessed using Bland-Altman analysis during the surgical and recovery phase.
Results:
A total of 271.8 hours of data in 99 patients was recorded simultaneously by the OHRM and patient monitor. The median coverage was 86% (IQR 65%-95%) and did not differ significantly between surgery and recovery (Wilcoxon paired difference test P=.17). Agreement analysis showed the limits of agreement (LoA) of the difference between the OHRM and the ECG HR were within the range of 5 beats per minute (bpm). The mean bias was -0.14 bpm (LoA between -3.08 bpm and 2.79 bpm) and -0.19% (LoA between -5 bpm to 5 bpm) for the PPG- measured HR compared to the ECG-measured HR during surgery; during recovery, it was -0.11 bpm (LoA between -2.79 bpm and 2.59 bpm) and -0.15% (LoA between -3.92% and 3.64%).
Conclusions:
This study shows that an OHRM equipped with a PPG sensor can measure HR within the ECG reference standard of -5 bpm to 5 bpm or -10% to 10% in the perioperative setting when the PPG signal is of sufficient quality. This implies that an OHRM can be considered clinically acceptable for HR monitoring in low acuity hospitalized patients.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pre-Procedural Guidelines for Assessing Blood Pressure