Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
Published on: May 19, 2020
Validation of a rapid liquid chromatography tandem mass spectrometric method for the quantitative analysis of
Susan C Scott1, Nicole M Anders1, Morgan Scardina1
1Sidney Kimmel Comprehensive Cancer Center, Johns Hopkins School of Medicine, Baltimore, MD, USA.
Abstract:
Vistusertib is an orally bioavailable mTOR inhibitor that is being studied in clinical trials. A novel reliable method was developed to quantitate vistusertib using LC-MS/MS to explore drug exposure-response relationships. Sample preparation involved protein precipitation using acetonitrile. Separation of vistusertib and the internal standard, AZD8055, was achieved with a Waters Acquity UPLC BEH C18 column utilizing isocratic elution over a 3 min total analytical run time. A SCIEX 4500 triple quadrupole mass spectrometer operated in positive electrospray ionization mode was used for the detection of vistusertib. The assay range was 5-5000 ng/mL and proved to be accurate (98.7-105.7%) and precise (CV ≤ 10.5%). A 40,000 ng/mL sample that was diluted 1:10 (v/v) with plasma was accurately quantitated. Long-term frozen plasma stability for vistusertib at -70 °C has been determined for at least 29 months. The method was applied for the measurement of plasma concentrations of vistusertib in a patient a solid tumor receiving 35 mg twice daily dose orally.
More Related Videos
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013