Semi-quantitative detection of inflammatory biomarkers using a laser-patterned multiplexed lateral flow device.
Alice H Iles1, Peijun J W He1, Ioannis N Katis1
1Optoelectronics Research Centre, University of Southampton, SO17 1BJ, UK.
Talanta
|November 5, 2021
Summary
This study introduces a novel multiplexed lateral flow device (LFD) for rapid, semi-quantitative detection of C-reactive protein (CRP) and procalcitonin (PCT). This advanced assay aids in faster patient triage and guides antibiotic treatment decisions for suspected pneumonia.
Area of Science:
- Biomarker detection
- Medical diagnostics
- Point-of-care testing
Background:
- C-reactive protein (CRP) and procalcitonin (PCT) are key inflammatory markers for infection diagnosis.
- Rapid and accurate detection of CRP and PCT can improve patient triage and antibiotic stewardship.
- Current methods may lack the speed or multiplexing capability for efficient near-patient assessment.
Purpose of the Study:
- To develop a novel multiplexed lateral flow device (LFD) for simultaneous, semi-quantitative detection of CRP and PCT.
- To enable rapid, near-patient assessment of inflammatory markers to guide clinical decisions.
- To demonstrate the feasibility of using laser-based partitioning for creating multi-parallel flow-paths in LFDs.
Main Methods:
- A standard LFD was modified using laser-based partitioning to create multiple parallel flow-paths.
- Each flow-path was engineered for specific detection of CRP (20-100 μg/mL) and PCT (>0.5 ng/mL).
- The multiplexed LFD was tested for semi-quantitative detection of CRP and PCT in spiked human serum samples.
Main Results:
- Successful semi-quantitative detection of both CRP and PCT was achieved using the multiplexed LFD.
- The device demonstrated tailored detection capabilities for each biomarker within defined ranges.
- The assay proved effective in spiked human serum, validating its performance.
Conclusions:
- The developed multiplexed LFD offers a rapid, near-patient solution for assessing key inflammatory markers.