Related Experiment Video
Updated: Oct 14, 2025

08:14
In Vitro Biochemical Assays using Biotin Labels to Study Protein-Nucleic Acid Interactions
Published on: July 17, 2019
12.9K
Serum biotin interference: A troublemaker in hormone immunoassays
Ümmühan Öncül1, Fatma Tuba Eminoğlu1, Engin Köse1
1Department of Pediatric Metabolism, Ankara University Faculty of Medicine, Ankara, Turkey.
Clinical Biochemistry
|November 5, 2021
Summary
High biotin levels can interfere with Roche assay results for 25-hydroxy vitamin D (25-OHD), parathyroid hormone (PTH), and thyroid-stimulating hormone (TSH). Streptavidin-coated magnetic particles (SMP) neutralization effectively removes this biotin interference.
Area of Science:
- Clinical Chemistry
- Biochemistry
- Endocrinology
Background:
- Biotin therapy is known to interfere with various immunoassay results.
- Accurate measurement of 25-hydroxy vitamin D (25-OHD), parathyroid hormone (PTH), and thyroid-stimulating hormone (TSH) is crucial for patient diagnosis and management.
Purpose of the Study:
- To investigate the interference of biotin on 25-OHD, PTH, and TSH assays.
- To evaluate the effectiveness of biotin neutralization using Streptavidin-coated magnetic particles (SMP).
Main Methods:
- Fifty children with biotinidase deficiency (BTD) undergoing biotin treatment were analyzed.
- Assays for 25-OHD, PTH, and TSH were performed using Roche Diagnostics and Beckman Coulter systems, alongside HPLC and LC/MS-MS.
- Biotin neutralization was achieved using SMP before and after sample analysis.
Main Results:
- Biotin interference was observed in Roche assays for 25-OHD, PTH, and TSH.
- Significant differences in results were noted after biotin neutralization with the Roche assay for 25-OHD (decrease) and PTH (increase).
- TSH levels also increased significantly in the Roche assay after neutralization.
Conclusions:
- High biotin concentrations significantly affect Roche assay results for PTH, 25-OHD, and TSH, posing a risk of misdiagnosis.
- SMP-based biotin neutralization effectively eliminates interference in these immunoassays.
- Clinical laboratories should consider biotin interference when interpreting results from specific assay systems.