Accuracy of vertebral puncture in percutaneous vertebroplasty
Tomoyuki Noguchi1,2,3,4, Koji Yamashita5, Yoshitaka Shida6
1Department of Radiology, National Hospital Organization Kyushu Medical Center, 1-8-1 Jigyohama, Chuo-ku, Fukuoka City, Fukuoka Province, 810-8563, Japan. tnogucci@gmail.com.
Purpose:
To clarify the accuracy of vertebral puncture of the vertebral tertile area needling (VETERAN) method puncturing the pedicle superimposed on one-third of the width between the lateral vertebral line to the contralateral medial lamina line compared with Cathelin-needle-assisted puncture (CAP) method puncturing using the Cathelin needle as a guide in percutaneous vertebroplasty.
Materials And Methods:
449 punctures by CAP method and 125 punctures by VETERAN method were enrolled. We compared the puncture accuracy of both methods. We estimated a vertebral estimated tilting ratio (VET-ratio) defined as ratio of the distance between the lateral vertebral line and the contralateral medial laminal line to the distance between the vertebral lateral line and the puncture point measured by computed tomography. We also estimated the procedural items and clinical outcomes.
Results:
VETERAN method with 100% of punctures within safe zone (cortical breaches within 2 mm) had significantly higher accuracy than CAP method with 97.8% (p < 0.01) for the 2 mm incremental evaluation. No cases with a VET-ratio of 36% or less had cortical breaches. VETERAN method had shorter operative time per puncture (p < 0.01) and exposure time per puncture (p < 0.05).
Conclusion:
VETERAN method reduced the occurrence of the inaccurate puncture, operative times, and exposure times. A VET-ratio with 36% or less is associated with a safe puncture using VETERAN method.