Repeatability and Reproducibility of Computed Tomography Radiomics for Pulmonary Nodules: A Multicenter Phantom Study
Xueqing Peng1, Shuyi Yang, Lingxiao Zhou2
1From the Institutes of Biomedical Sciences, Fudan University, Shanghai.
Background:
Radiomics can yield minable information from medical images, which can facilitate computer-aided diagnosis. However, the lack of repeatability and reproducibility of radiomic features (RFs) may hinder their generalizability in clinical applications.
Objectives:
The aims of this study were to explore 3 main sources of variability in RFs, investigate their influencing magnitudes and patterns, and identify a subset of robust RFs for further studies.
Materials And Methods:
A chest phantom with nodules was scanned with different computed tomography (CT) scanners repeatedly with varying acquisition and reconstruction parameters (April-May 2019) to evaluate 3 sources of variability: test-retest, inter-CT, and intra-CT protocol variability. The robustness of the RFs was measured using the concordance correlation coefficient, dynamic range, and intraclass correlation coefficient (ICC). The influencing magnitudes and patterns were analyzed using the Friedman test and Spearman rank correlation coefficient. Stable and informative RFs were selected, and their redundancy was eliminated using hierarchical clustering. Clinical validation was also performed to verify the clinical effectiveness and potential enhancement of the generalizability of radiomics research.
Results:
A total of 1295 RFs that showed all 3 sources of variability were included. The reconstruction kernel and the iteration level showed the greatest (ICC, 0.35 ± 0.31) and the least (ICC, 0.63 ± 0.27) influence on magnitudes. The different sources of variability showed relatively consistent patterns of influence (false discovery rate <0.001). Finally, we obtained a subset of 19 stable, informative, and nonredundant RFs under all 3 sources of variability. These RFs exhibited clinical effectiveness and showed better prediction performance than unstable RFs in the validation dataset (P = 0.017, Delong test).
Conclusions:
The stability of RFs was affected to different degrees by test-retest and differences in CT manufacturers and models and CT acquisition and reconstruction parameters, but the influences of these factors showed relatively consistent patterns. We also obtained a subset of 19 stable, informative, and nonredundant RFs that should be preferably used to enhance the generalizability of further radiomics research.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...