3D chest tomosynthesis using a stationary flat panel source array and a stationary detector: a Monte Carlo proof of
Thomas G Primidis1,2, Stephen G Wells3, Vadim Y Soloviev3
1Department of Physics, University of Liverpool, Liverpool, United Kingdom.
Abstract:
3D imaging modalities such as computed tomography and digital tomosynthesis typically scan the patient from different angles with a lengthy mechanical movement of a single x-ray tube. Therefore, millions of 3D scans per year require expensive mechanisms to support a heavy x-ray source and have to compensate for machine vibrations and patient movements. However, recent developments in cold-cathode field emission technology allow the creation of compact, stationary arrays of emitters. Adaptix Ltd has developed a novel, low-cost, square array of such emitters and demonstrated 3D digital tomosynthesis of human extremities and small animals. The use of cold-cathode field emitters also makes the system compact and lightweight. This paper presents Monte Carlo simulations of a concept upgrade of the Adaptix system from the current 60 kVp to 90 kVp and 120 kVp which are better suited for chest imaging. Between 90 kVp and 120 kVp, 3D image quality appears insensitive to voltage and at 90 kVp the photon yield is reduced by 40%-50% while effective dose declines by 14%. A square array of emitters can adequately illuminate a subject for tomosynthesis from a shorter source-to-image distance, thereby reducing the required input power, and offsetting the 28%-50% more input power that is required for operation at 90 kVp. This modelling suggests that lightweight, stationary cold-cathode x-ray source arrays could be used for chest tomosynthesis at a lower voltage, with less dose and without sacrificing image quality. This will reduce weight, size and cost, enabling 3D imaging to be brought to the bedside.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography