Efficacy and safety of a parylene-coated occluder for atrial septal defect: a prospective, multi-center, randomized
Kai Yang1,2, Jian-Hua Lyu1, Hai-Bo Hu1
1Center of Structural Heart Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing 100037, China.
Background:
Nitinol-containing devices are widely used in clinical practice. However, there are concerns about nickel release after nitinol-containing device implantation. This study aimed to compare the efficacy and safety of a parylene-coated occluder vs. a traditional nitinol-containing device for atrial septal defect (ASD).
Methods:
One-hundred-and-eight patients with ASD were prospectively enrolled and randomly assigned to either the trial group to receive a parylene-coated occluder (n = 54) or the control group to receive a traditional occluder (n = 54). The plugging success rate at 6 months after device implantation and the pre- and post-implantation serum nickel levels were compared between the two groups. A non-inferiority design was used to prove that the therapeutic effect of the parylene-coated device was non-inferior to that of the traditional device. The Cochran-Mantel-Haenszel chi-squared test with adjustment for central effects was used for the comparison between groups.
Results:
At 6 months after implantation, successful ASD closure was achieved in 52 of 53 patients (98.11%) in both the trial and control groups (95% confidence interval (CI): [-4.90, 5.16]) based on per-protocol set analysis. The absolute value of the lower limit of the 95% CI was 4.90%, which was less than the specified non-inferiority margin of 8%. No deaths or severe complications occurred during 6 months of follow-up. The serum nickel levels were significantly increased at 2 weeks and reached the maximum value at 1 month after implantation in the control group (P < 0.05 vs. baseline). In the trial group, there was no significant difference in the serum nickel level before vs. after device implantation (P > 0.05).
Conclusions:
The efficacy of a parylene-coated ASD occluder is non-inferior to that of a traditional uncoated ASD occluder. The parylene-coated occluder prevents nickel release after device implantation and may be an alternative for ASD, especially in patients with a nickel allergy.
More Related Videos
07:41Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
Related Concept Videos
Clinical Trials: Overview
Acute Coronary Syndrome III: Diagnostic Studies