A novel 5-ring multifocal electroretinography stimulus for detecting hydroxychloroquine retinal toxicity
Adrian Tsang1, Pushpinder Kanda2, Chloe Gottlieb1,3
1Department of Ophthalmology, The University of Ottawa Eye Institute, Ottawa, ON, Canada.
Documenta Ophthalmologica. Advances in Ophthalmology
|November 11, 2021
Summary
A new 5-ring multifocal electroretinogram (mfERG) protocol significantly reduces screening time for hydroxychloroquine (HCQ) retinopathy. This rapid mfERG method shows comparable results to standard protocols, improving clinical practicality for HCQ toxicity detection.
Area of Science:
- Ophthalmology
- Neuroscience
- Toxicology
Background:
- Multifocal electroretinogram (mfERG) is valuable for detecting early hydroxychloroquine (HCQ) retinopathy.
- Current mfERG protocols are limited by accessibility and long testing times, hindering widespread clinical adoption.
- HCQ retinopathy screening is crucial due to the drug's potential ocular toxicity.
Purpose of the Study:
- To evaluate a novel concentric 5-ring mfERG stimulus for a simplified and rapid HCQ toxicity screening protocol.
- To compare the efficacy and time efficiency of the 5-ring mfERG protocol against the standard 61-hexagon mfERG stimulus.
Main Methods:
- An observational cross-sectional study involving 52 patients undergoing HCQ retinopathy screening.
- Comparison of a novel 5-ring mfERG protocol with a standard 61-hexagon mfERG protocol.
- Assessment of test completion time and correlation of mfERG ring ratios with HCQ dosage and duration.
Main Results:
- The 5-ring mfERG protocol was significantly faster (1.3 min) than the 61-hexagon protocol (5.2 min).
- The R2/R5 ring ratio from the 5-ring protocol showed moderate correlations with daily, cumulative HCQ dose, and therapy duration.
- The 5-ring protocol's R2/R5 ring ratio demonstrated stronger correlations with daily and cumulative HCQ doses compared to the 61-hexagon stimulus.
Conclusions:
- The novel 5-ring mfERG protocol significantly reduces data acquisition time for HCQ retinopathy screening.
- This rapid mfERG protocol provides comparable results to the standard 61-hexagon stimulus for detecting HCQ-induced electrophysiological changes.
- The 5-ring mfERG protocol offers potential for increased clinical practicality in routine screening for HCQ toxicity.
Keywords:
Chloroquine retinopathyHydroxychloroquine retinopathyMultifocal electroretinogramRing ratio