Model-based approach to investigate equipment-induced error in pressure-waveform derived hemodynamic measurements
Masoud Farahmand1, Hossein Mirinejad2, Christopher G Scully1
1Division of Biomedical Physics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, United States of America.
None:
Objective.Advanced hemodynamic monitoring systems have provided less invasive methods for estimating pressure-derived measurements such as pressure-derived cardiac output (CO) measurements. These devices apply algorithms to arterial pressure waveforms recorded via pressure recording components that transmit the pressure signal to a pressure monitor. While standards have been developed for pressure monitoring equipment, it is unclear how the equipment-induced error can affect secondary measurements from pressure waveforms. We propose an approach for modelling different components of a pressure monitoring system and use this model-based approach to investigate the effect of different pressure recording configurations on pressure-derived hemodynamic measurements.Approach.The proposed model-based approach is a three step process. (1) Modelling the response of pressure recording components using bench tests; (2) verifying the identified models through nonparametric equivalence tests; and (3) assessing the effects of pressure recording components on pressure-derived measurements. To delineate the application of this approach, we performed a series of model-based analyses to quantify the combined effect of a wide range of tubing configurations with various damping ratios and natural frequencies and monitors with different bandwidths on pressure waveforms and CO measurements by six pulse contour algorithms.Results.Model-based results show the error in pressure-derived CO measurements because of tubing configurations with different natural frequencies and damping ratios. Tubing configurations with low natural frequencies (<23 Hz) altered characteristics of pressure waveforms in a way that affected the CO measurement, some by as much as 20%.Significance.Our method can serve as a tool to quantify the performance of pressure recording systems with different dynamic properties. This approach can be applied to investigate the effects of physiologic signal recording configurations on various pressure-derived hemodynamic measurements.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Typical Model Studies