Related Experiment Video
Updated: Oct 13, 2025

Implantation of Electroencephalogram and Electrocardiogram Telemetry Devices in Neonatal Rabbit Kits
Published on: February 28, 2025
ICD shocks and complications in patients with inherited arrhythmia syndromes
Matthew Siskin1, Marina Cerrone1, Mohamed Shokr1
1Leon H. Charney Division of Cardiology, NYU Langone Health, New York, NY, USA.
Background:
There is limited information on the long-term outcomes of ICDs in patients with inherited arrhythmia syndromes.
Methods:
Prospective registry study of inherited arrhythmia patients with an ICD. Incidence of therapies and complications were measured as 5-year cumulative incidence proportions and analyzed with the Kaplan-Meier method. Incidence was compared by device indication, diagnosis type and device type. Cox-regression analysis was used to identify predictors of appropriate shock and device complication.
Results:
123 patients with a mean follow up of 6.4 ± 4.8 years were included. The incidence of first appropriate shock was 56.52% vs 24.44%, p < 0.05 for cardiomyopathy and channelopathy patients, despite similar ejection fraction (61% vs 60%, p = 0.6). The incidence of first inappropriate shock was 13.46% vs 56.25%, p < 0.01 for single vs. multi-lead devices. The incidence of first lead complication was higher for multi-lead vs. single lead devices, 43.75% vs. 17.31%, p = 0.04. Patients with an ICD for secondary prevention were more likely to receive an appropriate shock than those with primary prevention indication (HR 2.21, CI 1.07-4.56, p = 0.03). Multi-lead devices were associated with higher risk of inappropriate shock (HR 3.99, CI 1.27-12.52, p = 0.02), with similar appropriate shock risk compared to single lead devices. In 26.5% of patients with dual chamber devices, atrial sensing or pacing was not utilized.
Conclusion:
The rate of appropriate therapies and ICD complications in patients with inherited arrhythmia is high, particularly in cardiomyopathies with multi-lead devices. Risk-benefit ratio should be carefully considered when assessing the indication and type of device in this population.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias I: Introduction