Effects of 1 MHz Therapeutic Ultrasound on Limb Blood Flow and Microvascular Reactivity: A Randomized Pilot Trial
Megan Waters1, Branko Miljkovic1, Jozelyn Rascon1,2
1Doctor of Physical Therapy Program, College of Health Sciences, The University of Texas at El Paso, El Paso, TX 79968, USA.
Abstract:
A randomized, double-blind, placebo-controlled, cross-over study where continuous therapeutic ultrasound (CUS; at 0.4 W/cm2), pulsed therapeutic ultrasound (PUS; at 20% duty cycle, 0.08 W/cm2), both at 1 MHz, and placebo (equipment on, no energy provided) were randomized and applied over the forearm of the non-dominant arm for 5 min in 10 young, healthy individuals. Absolute and peak forearm blood flow (FBF) were measured via Venous Occlusion Plethysmography. FBF was measured before, halfway, and after (immediately and 5 min after) the therapeutic ultrasound (TUS) intervention. Post-ischemic peak FBF was measured 10 min before and 10 min after the TUS intervention. A two-way repeated measures ANOVA (group × time) was selected to assess differences in FBF before, during, and after TUS treatment, and for peak FBF before and after TUS treatment. FBF increased 5 min after TUS in CUS compared to placebo (2.96 ± 1.04 vs. 2.09 ± 0.63 mL/min/100 mL of tissue, p < 0.05). PUS resulted in the greatest increase in Peak FBF at 10 min after US (Δ = 3.96 ± 2.02 mL/min/100 mL of tissue, p = 0.06). CUS at 1 MHz was an effective treatment modality for increasing FBF up to 5 min after intervention, but PUS resulted in the greatest increase in peak FBF at 10 min after intervention.