The Effect of Derotational Kirschner Wires on Fracture Gap Reduction With Variable-Pitch Headless Screws
Daniel Lynch1, John P Mickley1, Adam Gordon2
1Ohio State University College of Medicine, Columbus, OH.
Purpose:
We evaluated the impact of angled derotational Kirschner wires (K-wires) on fracture gap reduction with variable-pitch headless screws.
Methods:
Fully threaded variable-pitch headless screws (20 and 28 mm) were inserted into "normal" bone models of polyurethane blocks. In separate trials, derotational K-wires were inserted at predetermined angles of 0°, 15°, 30°, and 40° and compared with each other, with no K-wire as a control. Fluoroscopic images taken after each screw turn were analyzed. The optimal fracture gap closure, initial screw push-off, and screw back-out gap creation were determined and compared at various derotational K-wire angles.
Results:
Initial screw push-off due to screw insertion and screw back-out gap creation were not significantly affected by the angle of the derotational K-wire. With a 20-mm screw, only a 40° derotational K-wire led to significantly less gap closure compared with control and with 0°, 15°, and 30° derotational K-wires. It led to an approximately 60% decrease in gap closure compared with no K-wire. With the 28-mm screw, compared with no K-wire, 15° and 30° derotational K-wires led to statistically significant decreases in gap closure (approximately 25%), whereas a 40° derotational K-wire led to an approximately 60% decrease. With the 28-mm screw, the 40° derotational K-wire also led to a statistically significant smaller gap closure when compared with 0°, 15°, and 30° derotational K-wires.
Conclusions:
A derotational K-wire placed in parallel to the planned trajectory of a headless compression screw does not affect fracture gap closure. With greater angulation of the derotational K-wire, the fracture gap is still closed, but less tightly.
Clinical Relevance:
Derotational K-wires can help prevent fracture fragment rotation during headless compression screw insertion. At small deviations from parallel (≤30°), fracture gap closure achieved by the screw is minimally affected. At greater angles (ie, 40°), fracture gap closure may be substantially reduced, preventing fracture compression.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Screw: Problem Solving
To evaluate the...
Deformation in a Circular Shaft
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Plastic Deformation in Circular Shafts
Residual Stresses in Circular Shafts