Related Experiment Video
Updated: Oct 11, 2025

Techniques of Laparoscopic Right Posterior Sectionectomy: Glissonian Approach and a Parenchymal Transection Technique
Published on: October 6, 2023
[Design of Magnamosis System for Endoscopic Tissue Clamping]
Miaomiao Zhang1,2, Lin Ji1,2, Ting Lan2,3
1Department of Hepatobiliary Surgery, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061.
Abstract:
Based on the principle of magnetic anastomosis technique, the design of magnetic anastomosis system for endoscopic tissue clamping is proposed. The system includes a semi-ring magnet, a special structure transparent cap and a detachable push rod. With the help of the existing digestive endoscopy and endoscopic tissue gripper, the endoscopic close clamping and anastomosis of the bleeding or perforated tissue can be completed. After the anastomosis, the magnet falls off and is discharged through the digestive tract. Animal experiments showed that the system was easy to use, the fistula was clamped firmly, the magnet was discharged for 7~21 days, and there was no magnet retention and digestive tract obstruction. Further safety verification, optimization of endoscopic operation, the system can be used in clinical trial.