Spatially resolved free-induction decay spectroscopy using a 3D ultra-short echo time multi-echo imaging sequence
Anja Fischer1, Petros Martirosian1, Thomas Benkert2
1Section on Experimental Radiology, University Hospital Tübingen, Tübingen, Germany.
Purpose:
Biologically interesting signals can exhibit fast transverse relaxation and frequency shifts compared to free water. For spectral assignment, a ultra-short echo time (UTE) imaging sequence was modified to provide pixel-wise free-induction decay (FID) acquisition.
Methods:
The UTE-FID approach presented relies on a multi-echo 3D spiral UTE sequence with six echoes per radiofrequency (RF) excitation (TEmin 0.05 ms, echo spacing 3 ms). A complex pixel-wise raw data set for FID spectroscopy is obtained by several multi-echo UTE measurements with systematic shifting of the readout by 0.25 or 0.5 ms, until the time domain is filled for 18 or 45 ms. B0 drifts are compensated by mapping and according phase correction. Autoregressive extrapolation of the signal is performed before Gaussian filtering. This method was applied to a phantom containing collagen-water solutions of different concentrations. To calculate the collagen content, a 19-peak collagen model was extracted from a non-selective FID spectrum (50% collagen solution). Proton-density-collagen-fraction (PDCF) was calculated for 10 collagen solutions (2%-50%). Furthermore, an in vivo UTE-FID spectrum of adipose tissue was recorded.
Results:
UTE-FID signal patterns agreed well with the non-spatially selective pulse-acquire FID spectrum from a sphere filled with 50% collagen. Differentiation of collagen solution from distilled water in the PDCF map was possible from 4% collagen concentration for a UTE-FID sequence with 128 × 128 × 64 matrix (voxel size 1 × 1 × 2.85 mm3 ). The mean values of the PDCF correlate linearly with collagen concentration.
Conclusion:
The presented UTE-FID approach allows pixel-wise raw data acquisition similar to non-spatially selective pulse-acquire spectroscopy. Spatially resolved applications for assessment of spectra of rapidly decaying signals seem feasible.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...