Blood Flow Imaging with Ultrafast Doppler
Published on: October 14, 2020
Improving image contrast and accuracy in velocity estimation by convolution filters for intracardiac blood flow
Michiya Mozumi1, Ryo Nagaoka2, Hideyuki Hasegawa2
1Graduate School of Science and Engineering, University of Toyama, Toyama 930-8555, Japan.
Abstract:
In this study, the point spread function (PSF) of an ultrasound imaging system was estimated and used as a reference signal in a filtering method for improvement of image quality. The PSF of the imaging system was estimated from measured echo signals from an imaging target. Convolution filters (including deconvolution) were used for improvement of image contrast and spatial resolution. Furthermore, the accuracy in estimation of velocity vectors was evaluated for investigation of the impact of the proposed filters on velocity estimation. In the phantom experiment, contrast of the B-mode image was improved from 76.4 dB to 81.1 dB and 77.8 dB using the convolution and deconvolution filters, respectively. Also, the two-dimensional (2D) velocity distribution in the phantom was estimated by the block matching method, and the bias error (BE) in the estimated lateral velocity was reduced from -19.7% to 2.16% and 2.29% using the convolution and the deconvolution filters, respectively.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...