Abdominal aorta measurements by a handheld ultrasound device compared with a conventional cart-based ultrasound
Abdulrahman M Alfuraih1, Abdulaziz I Alrashed2, Saleh O Almazyad2
1From the Department of Radiology and Medical Imaging, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Alkharj, Saudi Arabia.
Background:
Ultraportable or pocket handheld ultrasound devices (HUD) may be useful for large-scale abdominal aortic aneurysm screening. However, the reproducibility of measurements has not been compared with conventional cart-based ultrasound machines.
Objectives:
Investigate the intra- and inter-operator reproducibility of a HUD compared with a conventional ultrasound machine for aortic screening.
Design:
Analytical, cross-sectional.
Setting:
Ultrasound department at a large tertiary care hospital in Riyadh.
Patients And Methods:
Eligible male participants aged ≥60 years were invited to participate upon arriving for a non-vascular ultrasound appointment. Three repeated anteroposterior measurements of the transverse aorta were made at the proximal and distal locations for each machine before repeating the measurements on a subset of participants by a second blinded operator. Intraclass correlation coefficients (ICC) and the Bland-Altman method were used to analyze reproducibility.
Main Outcome Measure:
Inter-system and intra- and inter-operator ICCs.
Sample Size:
114 males with repeated measurements by second operator on a subset of 35 participants.
Results:
The median age (interquartile range) of participants was 68 years (62-74 years). The intra- and inter-operator ICCs were all >0.800 showing almost perfect agreement except for the inter-operator reproducibility at the proximal location using a conventional machine (ICC= 0.583, P=.007) and the Butterfly device (ICC=0.467, P=.037). The inter-system ICCs (95% CI) were 0.818 (0.736-0.874) and 0.879 (0.799-0.924) at the proximal and distal locations, respectively. The mean difference in aortic measurement between the ultrasound systems was 0.3 mm (1.7%) in the proximal location and 0.6 mm (3.6%) in the distal location. In total, >91% of the difference in measurements between the machines was <3 mm. The mean scanning time was 4:16 minutes for the conventional system and 3:53 minutes for the HUD (P=.34).
Conclusions:
Abdominal aortic screening using a HUD was feasible and reliable compared with a conventional ultrasound machine. A pocket HUD should be considered for large-scale screening.
Limitations:
No cases of abdominal aortic aneurysm in the sample and lack of blinding.
Conflict Of Interest:
None.
More Related Videos
Related Concept Videos
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Abdominal Aorta
The celiac trunk, a singular artery, divides into the left gastric artery, which...
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection....
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: