Related Experiment Video
Updated: Oct 10, 2025

Making MR Imaging Child's Play - Pediatric Neuroimaging Protocol, Guidelines and Procedure
Published on: July 30, 2009
Simulation of SAR Induced Heating in Infants undergoing 1.5 T Magnetic Resonance Imaging
Abstract:
RF absorption in patients undergoing MRI procedures poses a major safety risk due to resulting heating in the tissue. In order to stay below permitted temperature limits the SAR has to be quantified and limited. Based on the model of an infant inside a birdcage coil we have investigated the SAR distribution in the body at 1.5T. Thermal simulations could thus be performed to establish a relationship between the limitations of SAR and temperature. Results show a thermal hotspot in the neck region caused by high local absorption. The temperature limits in this local area were exceeded after 7min of excitation within regulatory SAR limits. For a long-term exposure critical organs in the body's core also undergo thermal stress beyond limitations. This indicates the need for constraints in regard to long MR procedures to consider the temporal aspect of heating.Clinical Relevance-This work establishes a relationship between SAR and temperature in infants undergoing MRI and shows potential risks of long-term procedures due to induced thermal stress.