A complication of ECMO cannula placement resulting in hemodynamic and oxygenation alterations: A case report
Sergio B Navarrete1, Anne R Hermon1, Megan P Kostibas1
1Johns Hopkins University Hospital, USA.
Abstract:
Veno-veno extracorporeal membrane oxygenation (VV ECMO) is used as a bridge to recovery in acute respiratory distress syndrome (ARDS) patients who have reversible lung failure. We present a complication of ECMO cannula placement/position resulting in hemodynamic and oxygenation alterations. These demonstrate principles related to the interaction of the VV ECMO circuit and patient cardio-pulmonary physiology. Consideration and comprehension of pulmonary shunt fraction, ECMO cannula recirculation ratio and ECMO blood flow to cardiac output (CO) ratio are central to continuous assessment and diagnosis of cardio-pulmonary changes encountered during management of VV ECMO.
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Hemodialysis II: Procedure and Complications
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: