Assessment of Calibration Models for Cuff-Less Blood Pressure Measurement After One Year of Aging
Objective:
Many calibration models for cuff-less blood pressure (BP) measurement must be periodically updated with cuff BP values to account for vascular aging. However, the time period required for these "cuff re-calibrations" is largely unknown. The impact of one year of aging on several calibration models was assessed.
Methods:
Ten humans (6 males, 57±18 years, 3 hypertensives) were studied during multiple recording sessions that occurred one year apart. In each session, electrocardiography (ECG), ear photoplethysmography (PPG), finger PPG, and toe PPG waveforms and manual cuff BP were recorded before and after slow breathing, mental arithmetic, cold pressor, and nitroglycerin. Linear models based on each PPG waveform, which were previously shown to offer value in predicting the intervention-induced BP changes in a larger subject cohort, were employed. The model coefficients were determined for each subject via one session, and the fully-defined, subject-specific calibration models were then evaluated in the corresponding subjects via the session one year later.
Results:
Only a linear model relating toe pulse arrival time (PAT) - time delay between ECG R-wave and toe PPG foot - to systolic BP (SBP) remained useful. After the year, this model changed little on average (root-mean-squared-error (RMSE) = 1.5 mmHg) and predicted the cuff BP values better than the average of the initial cuff BP values of the subject (RMSE = 9.6±0.8 mmHg vs. 12.7±1.0 mmHg; p < 0.05).
Conclusion:
These results suggest annual cuff recalibrations for the toe PAT-SBP model.
Significance:
Toe PAT may offer a practical recalibration period that fosters user adherence.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...