Satisfied quantitative value can be acquired by short-time bone SPECT/CT using a whole-body cadmium-zinc-telluride
Tomohiko Yamane1,2, Masafumi Takahashi3, Yohji Matsusaka4
1Division of Nuclear Medicine, Department of Radiology, Saitama Medical University Hospital, 38 Moro-Hongo, Moroyama, 350-0495, Japan. yamane_t@saitama-med.ac.jp.
Abstract:
The aim of this study was to evaluate the quantitative values of short-time scan (STS) of metastatic lesions compared with a standard scan (SS) when acquired by whole-body bone SPECT/CT with cadmium-zinc-telluride (CZT) detectors. We retrospectively reviewed 13 patients with bone metastases from prostate cancer, who underwent SPECT/CT performed on whole-body CZT gamma cameras. STSs were obtained using 75, 50, 25, 10, and 5% of the list-mode data for SS, respectively. Regions of interest (ROIs) were set on the increased uptake areas diagnosed as metastases. Intraclass correlation coefficients (ICCs) of standardized uptake values (SUVs) for the ROIs were calculated between the SS and each STS, and ICC ≥ 0.8 was set as a perfect correlation. Moreover, the repeatability coefficient (RC) was calculated, and RC ≤ 20% was defined as acceptable. A total of 152 metastatic lesions were included in the analysis. The ICCs between the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 0.999, 0.997, 0.994, 0.983, and 0.955, respectively. The RCs of the SS vs. 75%-STS, 50%-STS, 25%-STS, 10%-STS, and 5%-STS were 7.9, 12.4, 19.8, 30.8, and 41.3%, respectively. When evaluating the quality of CZT bone SPECT/CT acquired by a standard protocol, 25%-STS may provide adequate quantitative values.
More Related Videos
08:04Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging