Related Experiment Video
Updated: Oct 9, 2025

09:43
Preparation of Silica Nanoparticles Through Microwave-assisted Acid-catalysis
Published on: December 16, 2013
18.9K
TEM sample preparation of microsized LiMn2O4 powder using an ion slicer
Jung Sik Park1, Yoon-Jung Kang2, Sun Eui Choi3
1Product Application Support, JEOL Korea, Seoul, 05355, South Korea. jspark@jeol.co.kr.
Applied Microscopy
|December 23, 2021
Summary
This study presents a new method for preparing transmission electron microscopy (TEM) samples from lithium-ion battery powders. Ion slicer milling ensures structural integrity for high-resolution scanning TEM (STEM) analysis.
Area of Science:
- Materials Science
- Nanotechnology
- Electrochemistry
Background:
- Transmission electron microscopy (TEM) is crucial for analyzing battery materials.
- Microsized powders from lithium-ion batteries present challenges for TEM sample preparation.
- Artefacts can compromise the reliability of TEM analysis.
Purpose of the Study:
- To develop an effective method for preparing TEM samples from microsized lithium-ion battery powders.
- To maintain the intrinsic structure of battery materials during sample preparation.
- To enable high-resolution scanning TEM (STEM) analysis with aberration correction.
Main Methods:
- Utilized ion slicer milling for precise thinning of battery powder samples.
- Employed argon-ion milling techniques to optimize specimen quality.
- Applied Cs-corrected STEM for high-resolution imaging and crystal integrity investigations.
Main Results:
- Successfully prepared TEM-transparent specimens from microsized lithium-ion battery powders.
- Minimized structural distortion and artefacts during the sample preparation process.
- Achieved high-resolution STEM observations, revealing crystal integrity.
Conclusions:
- Ion slicer milling is an effective technique for preparing TEM samples of battery materials.
- The developed method ensures specimen integrity for advanced electron microscopy.
- This approach enhances the reliability of TEM analysis for lithium-ion battery research.
Keywords:
Backside-ion millingBroad argon-ion beam (BIB)High-resolution (HR) STEMIon slicerLithium-ion secondary batteryTEM sample preparationMore Related Videos
Related Concept Videos
Sample Preparation for Analysis: Advanced Techniques
502
Accurate analysis of complex samples often requires advanced preparation techniques to achieve reliable and reproducible results. Samples containing inorganic or organic materials can be challenging to dissolve or decompose effectively. Standard sample preparation methods include acid digestion, fusion, dry ashing, and wet digestion.
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
502
Preparation of Samples for Electron Microscopy
6.0K
To be visualized by an electron microscope, either transmission or scanning, biological samples need to be fixed (stabilized) so the electron beam does not destroy them and dried thoroughly (desiccated/dehydrated) so the vacuum does not affect them. Fixation needs to be done as quickly as possible because the sample properties will start changing as soon as it is removed from its natural environment. For example, in a tissue sample, the oxygen levels begin decreasing, causing an altered...
6.0K