Ultrasound Lesion Detectability as a Distance Between Probability Measures

Summary

Generalized contrast-to-noise ratio (gCNR) offers a robust measure of lesion detectability (LD) in ultrasound imaging. This histogram-based method proves superior to traditional CNR, especially with modern beamformers and dynamic range variations.

Related Concept Videos

Ultrasonography01:17

Ultrasonography

Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
6.5K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.3K
Echo01:06

Echo

The human ear cannot distinguish between two sources of sound if they happen to reach within a specific time interval, typically 0.1 seconds apart. More than this, and they are perceived as separate sources.
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
643