Digital manometry to measure cerebrospinal fluid pressure during lumbar puncture
Robert Heider1, Peter G Kranz1, Erin Hope Weant1
1Department of Radiology, 213852Duke University Medical Center, Durham, NC, USA.
Rationale And Objectives:
Accurate cerebrospinal fluid (CSF) pressure measurements are critical for diagnosis and treatment of pathologic processes involving the central nervous system. Measuring opening CSF pressure using an analog device takes several minutes, which can be burdensome in a busy practice. The purpose of this study was to compare accuracy of a digital pressure measurement device with analog manometry, the reference gold standard. Secondary purpose included an assessment of possible time savings.
Materials And Methods:
This study was a retrospective, cross-sectional investigation of 71 patients who underwent image-guided lumbar puncture (LP) with opening CSF pressure measurement at a single institution from June 2019 to September 2019. Exclusion criteria were examinations without complete data for both the digital and analog measurements or without recorded needle gauge. All included LPs and CSF pressures were measured with the patient in the left lateral decubitus position, legs extended. Acquired data included (1) digital and analog CSF pressures and (2) time required to measure CSF pressure.
Results:
A total of 56 procedures were analyzed in 55 patients. There was no significant difference in mean CSF pressures between devices: 22.5 cm H2O digitally vs 23.1 analog (p = .7). Use of the digital manometer resulted in a time savings of 6 min (438 s analog vs 78 s digital, p < .001).
Conclusion:
Cerebrospinal fluid pressure measurements obtained with digital manometry demonstrate comparable accuracy to the reference standard of analog manometry, with an average time savings of approximately 6 min per case.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Cerebrospinal Fluid
CSF Production
CSF is produced mainly in the choroid plexus, a network of capillaries and ependymal cells located within the ventricular system of the brain....