Internal calibration as an emerging approach for endogenous analyte quantification: Application to steroids
Gioele Visconti1, Eulalia Olesti2, Víctor González-Ruiz2
1School of Pharmaceutical Sciences, University of Geneva, CMU - Rue Michel-Servet 1, 1211, Geneva, Switzerland; Institute of Pharmaceutical Sciences of Western Switzerland, University of Geneva, CMU - Rue Michel-Servet 1, 1211, Geneva, Switzerland.
This study introduces an innovative internal calibration method using stable isotope-labeled standards for mass spectrometry. This approach simplifies quantification and improves accuracy for clinical steroid analysis.
Area of Science:
- Analytical Chemistry
- Biochemistry
- Clinical Diagnostics
Background:
- Triple quadrupole mass spectrometry is standard for quantification.
- Traditional calibration curves are labor-intensive and error-prone.
- A need exists for more efficient and accurate calibration methods.
Purpose of the Study:
- To develop and validate an innovative internal calibration (IC) approach.
- To utilize a one-standard calibration with stable isotope-labeled (SIL) standards.
- To assess the method's performance for quantifying endogenous steroids in human serum.
Main Methods:
- Developed a novel internal calibration (IC) method using stable isotope-labeled (SIL) standards.
- Evaluated analyte-to-SIL response factor (RF) stability, SIL purity, and instrumental reproducibility.
- Applied the method to six endogenous steroids and tested on quality control and human serum samples.
Main Results:
- Consistent response factors (RFs) were observed across different LC-MS platforms.
- The IC method demonstrated satisfactory precision (1.3%-12.4%) and trueness (77.5%-107.0%).
- Results from the IC method closely matched conventional external calibration, with minimal bias.
Conclusions:
- The developed internal calibration approach offers a robust and accurate alternative for mass spectrometry quantification.
- This method simplifies the calibration process, reducing time and potential analytical errors.
- Naturally occurring isotopes in SIL standards can further enhance accuracy for low-concentration analytes.