Evaluation of the clinical chemistry tests analytical performance with Sigma Metric by using different quality specifications - Comparison of analyser actual performance with manufacturer data
Related Experiment Video
Updated: Oct 8, 2025

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
1Bursa Public Health Laboratory, Bursa, Turkey.
Introduction:
The interest in quality management tools/methodologies is gradually increasing to ensure quality and accurate results in line with international standards in clinical laboratories. Six Sigma stands apart from other methodologies with its total quality management system approach. However, the lack of standardization in tolerance limits restricts the advantages for the process. Our study aimed both to evaluate the applicability of analytical quality goals with Roche Cobas c 702 analyser and to determine achievable goals specific to the analyser used.
Materials And Methods:
The study examined under two main headings as Sigmalaboratory and Sigmaanalyser. Sigmalaboratory was calculated using internal and external quality control data by using Roche Cobas c 702 analyser for 21 routine biochemistry parameters and, Sigmaanalyser calculation was based on the manufacturer data presented in the package inserts of the reagents used in our laboratory during the study. Sigma values were calculated with the six sigma formula.
Results:
Considering the total number of targets achieved, Sigmaanalyser performed best by meeting all CLIA goals, while Sigmalaboratory showed the lowest performance relative to biological variation (BV) desirable goals.
Conclusions:
The balance between the applicability and analytical assurance of "goal-setting models" should be well established. Even if the package insert data provided by the manufacturer were used in our study, it was observed that almost a quarter of the evaluated analytes failed to achieve even "acceptable" level performance according to BV-based goals. Therefore, "state-of-the-art" goals for the Six Sigma methodology are considered to be more reasonable, achievable, and compatible with today's technologies.
Related Concept Videos
Data Validation
Key parameters for method validation include:
Development of Analytical Methods
Statistical Analysis: Overview
One of the most commonly used statistical quantifiers is the mean, which is the ratio between the sum of the numerical values of all results and the...
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Significance Testing: Overview