Dual-Layer Spectral CTA for TAVI Planning Using a Split-Phase Protocol and Low-keV Virtual Monoenergetic Images:
David Mangold1, Janek Salatzki2, Johannes Riffel2
1Department of Diagnostic and Interventional Radiology, University Hospital Heidelberg, Germany.
Purpose:
Adaptation of computed tomography protocols for transcatheter aortic valve implantation (TAVI) planning is required when a first-generation dual-layer spectral CT scanner (DLCT) is used. The purpose of this study was to evaluate the objective image quality of aortic CT angiography (CTA) for TAVI planning using a split-phase technique with reconstruction of 40 keV virtual monoenergetic images (40 keV-VMI) obtained with a DLCT scanner. CT angiography obtained with a single-phase protocol of a conventional single-detector CT (SLCT) was used for comparison.
Materials And Methods:
75 CTA scans from DLCT were retrospectively compared to 75 CTA scans from SLCT. For DLCT, spiral CTA without ECG-synchronization was performed immediately after a retrospectively ECG-gated acquisition covering the heart and aortic arch. For SLCT, spiral CTA with retrospective ECG-gating was performed to capture the heart and the access route simultaneously in one scan. Objective image quality was compared at different levels of the arterial access route.
Results:
40 keV virtual monoenergetic images of DLCT showed a significantly higher mean vessel attenuation, SNR, and CNR at all levels of the arterial access route. With 40 keV-VMI of DLCT, the overall mean aortic attenuation of all six measured regions was 589.6 ± 243 HU compared to 492.7 ± 209 HU of SLCT (p < 0.01). A similar trend could be observed for SNR (23.6 ± 18 vs. 18.6 ± 9; p < 0.01) and CNR (21.1 ± 18 vs. 16.4 ± 8; p < 0.01). No deterioration was observed for vascular noise (27.8 ± 9 HU vs. 28.1 ± 8 HU; p = 0.599).
Conclusion:
Using a DLCT scanner with a split-phase protocol and 40 keV-VMI for TAVI planning, higher objective image quality can be obtained compared to a single-phase protocol of a conventional CT scanner.
Key Points:
· Adaption of TAVI planning CT protocols may be required when using a first-generation dual-layer CT scanner.. · Reconstruction of virtual monoenergetic images at 40 keV improves image quality.. · With a split-phase protocol, the radiation dose is lower compared to a single-phase ECG-gated CT acquisition..
Citation Format:
· Mangold D, Salatzki J, Riffel J et al. Dual-Layer Spectral CTA for TAVI Planning Using a Split-Phase Protocol and Low-keV Virtual Monoenergetic Images: Improved Image Quality in Comparison with Single-Phase Conventional CTA. Fortschr Röntgenstr 2022; 194: 652 - 659.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...