A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Tibial anterior compartment compressibility in healthy subject, measured using compression sonography
Helen Anwander1, Livia Büchel1, Fabian Krause1
1Department of Orthopaedic Surgery and Traumatology, Inselspital, Bern University Hospital, University of Bern, Switzerland.
Introduction:
Compression sonography has been introduced for non-invasive measurement of compartment compressibility and possible diagnostic tool for acute or chronic compartment syndrome in studies using human cadavers and animal models. To date, standard values in healthy subjects are not yet defined. The aim was to define standard compartment compressibility values in healthy human subjects and to assess the reliability of this measurement method.
Methods:
In 60 healthy volunteers, using ultrasound, the diameter of the tibial anterior compartment was measured while applying no pressure, 10mbar and 80mbar of external pressure. A pressure manometer on the ultrasound head was used to monitor the externally applied pressure. Compartment compressibility ratio (R0-80, respectively R10-80) was calculated as following: The delta of the compartment diameter with high and low external pressure, divided through the diameter with low external pressure. In 10 volunteers, two examinators conducted each two measurements to assess the reliability.
Results:
Mean compartment compressibility ratio R10-80 was 15.9% ±3.6 (range: 7.2 - 22.2). Mean compartment compressibility ratio R0-80 was 18.2% ±5.0 (3.0 - 32.1). There was no significant correlation with lower leg circumference, height, weight, BMI, gender, hours of sport per week and type of sport (e.g. weightlifting/ cardio). For R10-80, intraobserver ICC 2.1 was 0.89 for an experienced observer and 0.79 for a non-experienced observer. Interobserver ICC 2.1 was 0.78. For R0-80, intraobserver ICC 2.1 was 0.71 for the experienced and 0.56 for the unexperienced observer. Interobserver ICC 2.1 was 0.59.
Discussion:
In healthy volunteers between 18 and 50 years of age, mean compartment compressibility ratio R10-80 was 15.9% ±3.6, independent of demographic factors and sport activity. Application of 10mbar instead of 0mbar increased image quality. Subsequently, R10-80 showed lower standard deviation and both higher intraobserver and interobserver reliability than R0-80. Using R10-80, this measurement method is reliable with very high intra- and interobserver correlation.
More Related Videos
08:48Author Spotlight: Exploring the Complexities of Achilles Tendon Injuries — Research and Future Directions
Published on: October 27, 2023
05:18Collection of Skeletal Muscle Biopsies from the Superior Compartment of Human Musculus Tibialis Anterior for Mechanical Evaluation
Published on: September 27, 2020
Related Concept Videos
Muscles of the Leg that Move the Foot and Toes
Anterior Compartment
The anterior compartment includes muscles that contribute to the dorsiflexion of the foot. This compartment houses the tibialis anterior, extensor hallucis longus, and extensor digitorum longus muscles....
Pulse Assessment Sites
Bones of the Lower Limb: Tibia and Fibula