The missing link: Unlocking the power of cardiac rhythm monitoring device based QT interval detection
Antony F Chu1, Gautham Rajagopal2, Shantanu Sarkar2
1Warren Alpert School of Medicine, Brown University, Rhode Island Hospital, 593 Eddy St, APC Building, Division of Cardiology, Providence, Rhode Island, USA.
Background:
The QT interval is of high clinical value as QT prolongation can lead to Torsades de Pointes (TdP) and sudden cardiac death. Insertable cardiac monitors (ICMs) have the capability of detecting both absolute and relative changes in QT interval. In order to determine feasibility for long-term ICM based QT detection, we developed and validated an algorithm for continuous long-term QT monitoring in patients with ICM.
Methods:
The QT detection algorithm, intended for use in ICMs, is designed to detect T-waves and determine the beat-to-beat QT and QTc intervals. The algorithm was developed and validated using real-world ICM data. The performance of the algorithm was evaluated by comparing the algorithm detected QT interval with the manually annotated QT interval using Pearson's correlation coefficient and Bland Altman plot.
Results:
The QT detection algorithm was developed using 144 ICM ECG episodes from 46 patients and obtained a Pearson's coefficient of 0.89. The validation data set consisted of 136 ICM recorded ECG segments from 76 patients with unexplained syncope and 104 ICM recorded nightly ECG segments from 10 patients with diabetes and Long QT syndrome. The QT estimated by the algorithm was highly correlated with the truth data with a Pearson's coefficient of 0.93 (p < .001), with the mean difference between annotated and algorithm computed QT intervals of -7 ms.
Conclusions:
Long-term monitoring of QT intervals using ICM is feasible. Proof of concept development and validation of an ICM QT algorithm reveals a high degree of accuracy between algorithm and manually derived QT intervals.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias