Impact of ultra-conservative ICD programming in patients with LVADs: Avoiding potentially unnecessary tachy-therapies
Alexander Robinson1, Valay Parikh1, Mohammad-Ali Jazayeri1
1Department of Cardiovascular Medicine, The University of Kansas Health System, Kansas City, Kansas, USA.
Background:
Patients with left ventricular assist devices (LVAD) often tolerate ventricular arrhythmias (VA). We aim to assess the frequency and outcomes of ICD therapies averted by ultraconservative ICD programming (UCP) in LVAD patients.
Methods:
This single center, retrospective cohort study included patients with LVADs and ICDs implanted from 2015 to 2019 that had UCP. The aim for UCP was to maximally delay VA treatments and maximize anti-tachycardia pacing (ATP) prior to ICD shocks. VA events were reviewed after UCP and evaluated under prior conservative programming to assess for potentially averted events (that would have resulted in either ATP or defibrillation with prior programming).
Results:
Fifty patients were included in the study with follow-up of median 16 ± 10.2 months after UCP. The median time from LVAD implantation to reprogramming was 7 days (IQR 5-9 days). Fourteen patients (28%) had potentially averted VA events that would have been treated with their prior ICD programming (82 total events, median two events per patient, IQR 1-10 events). Treated VA events occurred in 15 patients (30%). Eleven of the 14 patients with potentially averted VAs had treated events as well. Only one patient reported definitive symptoms of self-limited "dizziness" during a potentially averted event that did not result in hospitalization. No patients died of complications from or needed emergent care/hospitalization due a potentially averted VA.
Conclusions:
UCP in LVAD patients likely prevented unnecessary VA treatments in many patients with minimal reported symptoms during these potentially averted events. Prospective studies are necessary to confirm these findings.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Heart Failure VI: Adjunct Therapies
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...